You are here:

Electrocardiograph / ECG1212G

Electrocardiograph / ECG1212G

Electrocardiograph / ECG1212G

Description:

Electrocardiograph Touch Screen

ECG1212G

The device is an electrocardiograph which can collect 12-lead ECG signal simultaneously and print ECG waveform with thermal printing system. It features in, recording and displaying ECG waveform in auto / manual mode; measuring, analyzing and diagnosing ECG waveform parameters automatically; prompting for “Lead off” and “Lack of paper”; optional interface language (Chinese / English, etc.); built-in lithium battery, AC / DC; selecting rhythm lead freely to observe abnormal heart rhythm; case database management.

Functions :

  1.  Adopt 10.1” high-resolution color LCD, operate by full-touch screen and function bu ttons, convenient to use.
  2.  Sync collection for 12-lead ECG, support 12-lead and Cabrera-lead waveform display. Get high-quality ECG waveform via power frequency filter (50 / 60 Hz), baseline filter and EMG filter (25 / 35 Hz) of ECG signal.
  3.  Display 3 / 6 / 12 – lead ECG waveform, HR value, print mode, sensitivity, paper speed, filter state, clock, battery power, background grid, measured parameter and interpretation, etc. on one screen, and with prompt functions for lead – off, overload and system working state.
  4.  Power supply: AC / DC (which can adapt AC 50 / 60Hz), with built-in rechargeable lithium battery, charging circuit, perfect over-current and over-voltage protection circuit.
  5.  In optimal DC state, up to 10-hour standby time, continuous print for more than 3-hour, which meets the requirements of visiting a patient at home and body examination.
Specification :  
Input mode : floating and defibrillation protection
Lead : standard 12-lead
Patient leak current : < 10 µA
Input impedance : ≥ 2.5 MΩ
Time constant : ≥ 3.2 s
CMRR : > 105 dB
Recording mode : thermal printing system
Paper size : 210 mm (W) × 20 m (L), high-speed thermal paper
Paper speed : 12.5 mm/s, 25 mm/s, 50 mm/s, error: ± 5 %
Gain (sensitivity) : 5, 10, 20 mm / mV, accuracy: ± 2 %
Standard sensitivity : 10 mm/mV ± 0.2 mm/mV
Record mode : auto, rhythm and manual record
Measurement parameters : HR, P – R interval, P Duration, QRS Duration, T Duration, Q – T interval, Q – Tc, P Axis, QRS Axis, T Axis, R (V5) amplitude, S (V1) amplitude, R (V5) + S (V1) amplitude
Safety class : class I, defibrillation-proof type CF applied part
Polarizing voltage : ± 610 mV
Noise level : ≤ 12 µVp-p
Sampling frequency for ECG signal input :  32 kHz
Sampling frequency for waveform data processing : 1 kHz
Pacemaker detection channel : II
Time deviation in each channel : <100 µs
Power supply :
  • AC: 100V ~ 240V (50 / 60 Hz)
  • DC: 14.8 V / 5200 mAh rechargeable lithium battery
Working mode : Continuous operation  

Categories: , ,

Description:

Categories: , ,